MEMORY HiCORDER 8860-50, 8861-50

8860-50

Call 02 9150 7744 For the BEST DEAL

MEMORY HiCORDER 8860-50

As an Oscilloscope, As a Data Logger
Record Waveforms in Any Situation
  • A faster CPU greatly enhances instrument operability and response.
  • Multi-channel logging on up to 64 or 128 channels (Use with the 8958)
  • 20 MS/s high-speed sampling (Use with the 8956)
  • Various interfaces built-in: LAN/USB/PC-card slot
  • Supports the new high-voltage input module for measuring high voltage directly.
The MEMORY HiCORDER 8860-50 and 8861-50 cannot operate alone. You must install one or more optional input modules in the unit.
Note: Memory boards are not built-in as a standard feature. Choose one board for Model 8860-50, and two of the same capacity for the 8861-50, for factory pre-installation.
Compare MEMORY HiCORDERs
 
 
Basic specifications (Accuracy guaranteed for 1 year)
  8860-50 8861-50
Number of input Units Max. 4 units Max. 8 units
Number of channels Max. 16 analog channels (max. 64 channels with scanner unit) + 16 logic channels (standard configuration) Max. 32 analog channels (max. 128 channels with scanner unit) + 16 logic channels (standard configuration)
Measurement ranges
(20 div full-scale)
5mV to 20V/div, 12 ranges, (using the 8956), resolution : 1/100 of range 5mV to 20V/div, 12 ranges, (using the 8956), resolution : 1/100 of range
Max. allowable input DC 400V (using the 8956) DC 400V (using the 8956)
Frequency characteristics DC to 10MHz (using the 8956) DC to 10MHz (using the 8956)
Power consumption 100 to 240 V AC (50/60 Hz) (220 VA max. printer not used)
12 V DC (use the DC POWER UNIT 9684 : option, factory installation only)
100 to 240 V AC (50/60 Hz) (280 VA max. printer not used)
12 V DC (use the DC POWER UNIT 9684 : option, factory installation only)
Dimensions, mass 330 mm (12.99 in) W × 250 mm (9.84 in) H × 184.5 mm (7.26 in) D, 8 kg (282.2 oz) (printer not installed) 330 mm (12.99 in) W × 250 mm (9.84 in) H × 284.5 mm (11.20 in) D, 10.5 kg (370.4 oz) (printer not installed)
Time axis
at memory function
5 µs to 5 min/div, 26 ranges, sampling period : 1/100 of range, external sampling, dual timebase possible
Measurement functions MEM (high-speed recording) REC (real-time recording) REC & MEM (realtime recording + high-speed recording) FFT (frequency analysis) Real-time Save (records directly to storage media)
Memory capacity 12-bits × 32M-Words/ch (1ch at 8860-50, 2ch at 8861-50) to 2M-Words/ch (16ch at 8860-50, 32ch at 8861-50) *Memory capacity can be expanded 32 times.(Optional memory board)
Data storage media PC Card Type II slot × 2, Hard disk drive (option) × 1
Recording paper (option) A4: 216 mm (8.50 in) × 30 m (98.43 ft), or A6: 112 mm (4.41 in) × 18 m (59.06 ft) selectable, thermal paper roll, Recording speed : Max. 25 mm (0.98 in)/sec
Display 10.4 inch TFT color LCD (SVGA, 800 × 600 dots)
External interfaces GP-IB (PC card), USB2.0 (series A receptacle 3 ports), LAN (HTTP/FTP/ Send mail or other), Monitor output (15-pin D-sub, SVGA output)
Supplied accessories Power cord × 1, Input cord label × 1, Application Disk (Wave Viewer Wv, Communication Commands table) × 1

 

MEMORY HiCORDER 8860-50 (8860-50E7-19M-00KR.pdf, 5,560 Kb) [Download]